Version 5.01 – July 2012 / SW Version: 2012.06.12Errore. Per applicare Nadpis 1 al testo da visualizzare in questo punto, utilizzare la scheda Home.
© DRC Srl 2012
46/102
4.6 SETTING Menu
After depressing this button, the instrument
checks some specific settings and, if
necessary,
adjusts
settings
to
make
possible automatic calibration.
This menu is used for automatic calibration
of PROBE DELAY and VELOCITY values
for specific probe (angle beam or dual
element) connected to the instrument.
♪
)))
Switches alarm ON/OFF and adjusts the alarm volume, as well as key
beep volume.
DISPLAY
Switches ON (
DISPLAY
) / OFF (
DISPLAY
) the display settings – see
4.7. DISPLAY Sub-Menu (SETTING):
SET_01x
Switches a memory group for saving/loading set-ups (SET_01x to
SET_09x) – see also 3.6, 3.7.
CURVES
Submenu for setting of evaluation curves parameters – see 4.8. CURVES
Sub-Menu (SETTING).
PROBE
Submenu for settings of specific parameters of transducers – see
4.9. PROBE Sub-Menu (SETTING).
A.CALIB
Switches ON/OFF the AUTOCALIBRATION mode – see 7.2. Automatic
Calibration of Probe Delay and Sound Velocity.
Settings of sound paths S1 and S2 for automatic calibration of PROBE
DELAY and VELOCITY values. Options are as follows:
-
S1 SET
setting of S1 sound path (just after activation of
A.CALIB
menu, depress AUTOCALIB left/yellow twice);
-
S2 SET
setting of S2 sound path (just after activation of
A.CALIB
menu, depress AUTOCALIB left/yellow once);
-
S1.S2 SET
setting of S1 and S2 sound paths simultaneously,
provided that S2 = 2 * S1 (is displayed just after
activation of
A.CALIB
menu);
-
S1 MEASURE
measuring of the echo position with the sound path S1;
-
S2 MEASURE
measuring of the echo position with the sound path S2;
-
CALC
calculation of PROBE DELAY and VELOCITY values.