Other Functions of GATE 1
-
Acoustic alarm may be switched ON, when the echo reaches or exceeds the
GATE 1 level or, with DAC switched ON, when the echo reaches or exceeds the
DAC level within the GATE 1 (GATE 1 level has no effect, when DAC is ON).
-
ON+AUX: analog output may be switched ON. There is a signal on the AOUT1
output (LEMO Analog Out), with voltage corresponding to relative echo position
within GATE 1 (see also 6. Functions of Analog Output):
minU = GATE1_BEGIN;
maxU = GATE1_END.
-
When GATE1 is set to ON+AUX and GATE2 is ON, there is a thickness
value on the AOUT2 output.
-
When GATE1 is set to ON+AUX and GATE2 is OFF, there is an echo
amplitude within the GATE1 on the AOUT2 output.
-
ON+A/TTL: sends echo position to AOUT1 ouptut and a TTL logic signal about
echo presence within GATE1 on AOUT2 output (see also 6. Functions of Analog
Output).
Note: If there is no echo within GATE 1, then maxU is on the output.