3. Keyboard Description
Display
Control Keys for Values Changing
Power
ON/OFF
Display Freeze
ON/OFF
Display
Brightness
Immediate
Execution
Keys
Escape Key
Save
Setup / Picture
Load
Setup / Picture
Quick Save
Picture
Zoom
Help
Lock
Status and
Calculation Window
3.1
Power Button
The POWER button
switches instrument power ON and OFF. Immediately after
power is turned ON, an initialization routine is started. During the start-up routine, the
instrument’s microprocessor measures and corrects circuit offsets. Also, the
microprocessor verifies that all circuits are operating properly.
After the “
Check and then press ‘Lock’ for pulser enable..
” message had been
displayed, make sure that the appropriate probe (transducer) is connected properly
and depress the
key in order to activate pulser.
NOTE: Long depressing the
button will switch OFF the instrument, without
saving last settings conditions.