UltiMate 3000 Series: Flow Managers and Column Thermostats
Operating Instructions
Page
91
10 Technical Appendix - Pin Assignment
Two 6-pin Mini-DIN ports on the instrument's rear panel (
→
Fig. 13, page 21) can be used in
Chromeleon to exchange digital signals with external instruments.
Fig. 47: Digital I/O (view from the rear)
* NO = normally open contact
If the relay is activated, the connection is between Com and NO.
Description
Digital Input
↔
Digital_GND
0 to +5V
Relay_Com
↔
Relay_NO
Switching voltage: 100 V
DC
; switching current: 0.25 A
Carry current: 0.5 A; Switching capacity: 3 W
Contact resistance: max. 150 m
Ω
Fig. 48: Digital I/O
The table below lists the functions assigned to the connector pins and the color and label of
the cable wire connected to each pin:
Pin
Signal Name
Signal Level
Core Color
Core Label
1
Not occupied
Pink
Analog High
2
Not occupied
Gray
Analog Low
3 Digital_GND 0V
Green
Digital
GND
5
Digital Input
0 to +5V
Yellow
Digital Input
6 Relay_NO
Brown
Digital
Output
8 Relay_Com
100V
DC
/0.25A/3W
White Dig./Out
(GND)
Black Shield
Fig. 49: Pin assignment (
6-pin Mini-DIN port and cable)
Relay_NO*
Digital_GND
Not used
Relay_Com
Digital Input
Not used
Summary of Contents for UltiMate FLM-3100
Page 2: ...UltiMate 3000 Series Flow Managers and Column Thermostats Operating Instructions...
Page 4: ...UltiMate 3000 Series Flow Managers and Column Thermostats Page II Operating Instructions...
Page 30: ...UltiMate 3000 Series Flow Managers and Column Thermostats Page 24 Operating Instructions...
Page 46: ...UltiMate 3000 Series Flow Managers and Column Thermostats Page 40 Operating Instructions...
Page 98: ...UltiMate 3000 Series Flow Managers and Column Thermostats Page 92 Operating Instructions...