If communication between Chromeleon and the flow manager cannot be established,
related error messages may appear in the Chromeleon audit trail.
Refer to the table below for the most important error messages, along with possible causes
and the corresponding remedial actions. (The number 1610103 in the table below is the
serial number of the instrument.)
Message
Probable Cause
Remedial Action
FLM-3x00@USB-1610103 -
Device not found on the USB.
The USB connection between the flow
manager and the Chromeleon server is
interrupted.
Check the USB connection.
The power supply to the flow manager
is interrupted.
Check the mains connection of the
flow manager.
Error opening FLM-3x00@USB-
1610103 – The System cannot find
the file specified
The USB connection between the flow
manager and the Chromeleon server is
interrupted.
Check the USB connection.
The power supply to the flow manager
is interrupted.
Check the mains connection of the
flow manager.
Error issuing control request to
FLM-3x00@USB-1610103
The Chromeleon server cannot connect
to the specified flow manager.
Check the USB connection.
Check the mains connection of the
flow manager.
Remove the flow manager specified in
the message from the server
configuration. Or else, select a
different flow manager from the list of
available flow managers in the server
configuration (via Properties/
Browse).
Error reading from FLM-3x00@
USB-1610103 Data error (cyclic
redundancy check)
There is a transmission error between
the flow manager and the Chromeleon
server.
Check the USB connection. The
connection to the next hub must not
exceed 5 m. The overall connection
length, including the hub connections
must not exceed 30 m (
→
Replace any defective USB cable or
hub.
Error reading from FLM-3x00@
USB-1610103
The connection between the flow
manager and the Chromeleon server is
interrupted.
Check the USB connection.
The power supply to the flow manager
is interrupted.
Check the mains connection of the
flow manager.
Summary of Contents for UltiMate FLM-3100
Page 2: ...UltiMate 3000 Series Flow Managers and Column Thermostats Operating Instructions...
Page 4: ...UltiMate 3000 Series Flow Managers and Column Thermostats Page II Operating Instructions...
Page 30: ...UltiMate 3000 Series Flow Managers and Column Thermostats Page 24 Operating Instructions...
Page 46: ...UltiMate 3000 Series Flow Managers and Column Thermostats Page 40 Operating Instructions...
Page 98: ...UltiMate 3000 Series Flow Managers and Column Thermostats Page 92 Operating Instructions...