Flush the splitter to remove any particles that may have polluted the splitter during the
exchange of the cartridge. To do so, place a lint-free cloth under the splitter to absorb
the solvent that leaks out (
→
Fig. 45: Place a cloth under the splitter
9.
Close the front panel.
10.
Run the pump for about 2 minutes.
11.
Open the front panel and remove the cloth.
12.
Reconnect the capillary. The splitter is now ready for operation.