UltiMate 3000 Series: Flow Managers and Column Thermostats
Page
64
Operating Instructions
Message
Probable Cause
Remedial Action
Left chip card
communication error!
The left chip card (slot A or B) is
not inserted properly or
defective.
Verify that the chip card is
installed properly
Left MSV communication
error!
The left switching may not be
connected properly or is
defective.
Turn the flow manager off and
on again
Left MSV controller error!
The left switching valve is
installed but cannot be accessed.
Retry
Left MSV error!
An error has been reported for
the left switching valve.
Contact Dionex Service if the
messages appears reapeatedly.
Left MSV error!
Retrying…
and/or
Left MSV position error!
The left switching valve cannot
be switched to the desired
position.
Switching is repeated
automatically. If this is not
successful, the
Left MSV
position error!
message
appears.
MSV communication error!
The swichting valve may not be
installed properly or is defective.
Turn the flow manager off and
on again. Contact Dionex
Service if the message appears
again.
Rear temperature sensor
communication error!
The temperature sensor for the
heat exchanger may not be
installed properly or is defective.
Turn the flow manager off and
on again.
Right chip card
communication error!
The left chip card (slot C or D) is
not inserted properly or
defective.
Verify that the chip card is
installed properly
Right MSV communication
error!
The right switching valve may
not be connected properly or is
defective.
Turn the flow manager off and
on again.
Right MSV controller error!
The right switching valve is
installed but cannot be accessed.
Retry.
Right MSV error!
An error has been reported for
the right switching valve.
Contact Dionex Service if the
messages appears reapeatedly.
Right MSV error!
Retrying….
and/or
Right MSV position error
The right switching valve cannot
be switched to the desired
position.
Switching is repeated
automatically. If this is not
successful, the
Right MSV
position error!
message
appears.
Summary of Contents for UltiMate FLM-3100
Page 2: ...UltiMate 3000 Series Flow Managers and Column Thermostats Operating Instructions...
Page 4: ...UltiMate 3000 Series Flow Managers and Column Thermostats Page II Operating Instructions...
Page 30: ...UltiMate 3000 Series Flow Managers and Column Thermostats Page 24 Operating Instructions...
Page 46: ...UltiMate 3000 Series Flow Managers and Column Thermostats Page 40 Operating Instructions...
Page 98: ...UltiMate 3000 Series Flow Managers and Column Thermostats Page 92 Operating Instructions...