UltiMate 3000 Series: Flow Managers and Column Thermostats
Page
58
Operating Instructions
5.5 Column Identification System (Column ID)
The flow manager is fitted with a column identification system (column ID) that allows
you to store column-specific information on a chip card and read it out whenever you
want. The column is shipped with the chip card attached.
1.
Insert the chip card in one of the four memory card slots (
→
the Dionex logo faces upward.
Fig. 39: Inserting a memory chip card in a memory card slot
Tips:
Column identification is supported for more than one column
simultaneously. When the chip card is installed correctly in the slot, the
slot LED is green.
As standard, the second slot from the right is intended for the splitter ID
chip card (
→
Splitter Identification System (Splitter ID), page 59).
Nevertheless, you may use this slot also for the column ID chip card if
you have specified this on the
Components
page (
→
page 53) in the
Server Configuration program.
2.
To store and read out the column-specific information, use the related
Column
commands and properties in the Chromeleon
Commands
dialog box. These entries
are available only if you have made the relevant settings in the flow manager's
properties (on the
Components
tab page in the Server Configuration program;
→
page 53). The stored information can then provide a GLP-compliant overview of
the column status.
Tip:
To open the
Commands
dialog box, open a Chromeleon control panel
and select
Command
on the
Control
menu or press the F8 key. In the
left list box, click the "+" sign beside
ColumnOven
to display the items
underneath.
For more information about the column identification system, refer to the
Chromeleon
online Help
.
Summary of Contents for UltiMate FLM-3100
Page 2: ...UltiMate 3000 Series Flow Managers and Column Thermostats Operating Instructions...
Page 4: ...UltiMate 3000 Series Flow Managers and Column Thermostats Page II Operating Instructions...
Page 30: ...UltiMate 3000 Series Flow Managers and Column Thermostats Page 24 Operating Instructions...
Page 46: ...UltiMate 3000 Series Flow Managers and Column Thermostats Page 40 Operating Instructions...
Page 98: ...UltiMate 3000 Series Flow Managers and Column Thermostats Page 92 Operating Instructions...