UltiMate 3000 Series: Flow Managers and Column Thermostats
Operating Instructions
Page
55
Option
Description
Temperature Signal
The
Temperature Signal
check box is selected by default. Accept this setting if
you want to record the temperature as a separate channel. (The channel records the
pressure at the splitter outlet.)
Accept the default name (
ColumnOven_Temp
) under which the temperature signal
is identified in the installation environment and in the Chromeleon client or enter a
new name. If you change the default device name, you have to re-link the
corresponding controls of the corresponding control panels.
Pressure Signal
The
Pressure Signal
check box is selected by default. Accept this setting if you
want to record the column pressure as a separate channel. Accept the default name
(
ColumnPressure
) under which the column pressure signal is identified in the
installation environment and in the Chromeleon client or enter a new name. If you
change the default device name, you have to re-link the corresponding controls of
the corresponding control panels.
Pressure Unit
Select the pressure unit to be used.
The following options are available: bar, MPa, and psi.
Flow Control
The
Flow Control
check box is selected by default. The additional channels that
are recorded and displayed under
Column Details
on the FLM control panel are
required when performing Operational Qualification or running a Flow Control Test
for the flow manager. In addition, some of the recorded values provide valuable
information to Dionex Service for troubleshooting in case the flow manager does
not pass Operational Qualification successfully.
Rear Temperature
The
Rear Temperature
check box is selected by default. The additional channels
that are recorded to monitor the temperature control in the column compartment and
at the rear of the instrument are required running a Thermo Unit Test for the flow
manager.
Relays Tab Page
The
Relays
tab page lists all available relays (FLM3x00_RELAY_1 and
FLM3x00_RELAY_2). Select a check box to enable or disable the corresponding relay. Make
sure that the required relays are selected; if they are not, they will not be available in
Chromeleon.
Inputs Tab Page
The
Inputs
tab page lists all available remote inputs (FLM3x00_INPUT_1 and
FLM3x00_INPUT_2). Select a check box to enable or disable the corresponding remote input.
Make sure that the required inputs are selected; if they are not, they will not be available in
Chromeleon.
Error Levels Tab Page
The
Error Levels
tab page classifies the severity of any errors that occur. It is generally not
necessary to change the default settings.
Tip:
Before turning off the flow manager by the power switch, always
"
disconnect
" the module in Chromeleon. Disconnecting the module in
Chromeleon is not required to set the flow manager to the Standby mode.
Summary of Contents for UltiMate FLM-3100
Page 2: ...UltiMate 3000 Series Flow Managers and Column Thermostats Operating Instructions...
Page 4: ...UltiMate 3000 Series Flow Managers and Column Thermostats Page II Operating Instructions...
Page 30: ...UltiMate 3000 Series Flow Managers and Column Thermostats Page 24 Operating Instructions...
Page 46: ...UltiMate 3000 Series Flow Managers and Column Thermostats Page 40 Operating Instructions...
Page 98: ...UltiMate 3000 Series Flow Managers and Column Thermostats Page 92 Operating Instructions...