5.3 Assigning the Flow Splitter
If the UltiMate 3000 system includes a flow manager and a pump (DGP-3600M,
LPG-3400M or HPG-3x00A/M), use the
Devices
tab page of the pump's properties to
indicate which pump is fluidically connected with the flow splitter:
1.
In the Chromeleon
Server Configuration
program, select the pump and click
Properties
on the pump's context menu.
2.
On the
Devices
tab page, specify which pump is connected with the flow splitter:
Fig. 38: Devices tab page (here: DGP-3600)
The default entry in the
uses FLM-3x00 Flow Splitter
field is
<None>
(no splitter). To
change the setting, click the arrow next to the input field and select a flow splitter from
the drop-down list.
Tips:
When you assign a flow splitter to the pump, the pump flow settings
apply directly to the flow on the splitter outlet, i.e., the split ratio is
considered automatically.
In the standard configuration of the UltiMate 3000 system with a
DGP-3600M, Dionex recommends assigning the flow splitter to the right
pump (MicroPump).
Summary of Contents for UltiMate FLM-3100
Page 2: ...UltiMate 3000 Series Flow Managers and Column Thermostats Operating Instructions...
Page 4: ...UltiMate 3000 Series Flow Managers and Column Thermostats Page II Operating Instructions...
Page 30: ...UltiMate 3000 Series Flow Managers and Column Thermostats Page 24 Operating Instructions...
Page 46: ...UltiMate 3000 Series Flow Managers and Column Thermostats Page 40 Operating Instructions...
Page 98: ...UltiMate 3000 Series Flow Managers and Column Thermostats Page 92 Operating Instructions...