94
Technical Reference
Moisture Meter User Manual 4.0
Memory Failure,
Readings Corrupt,
Settings Corrupt
See Readings can't be stored on page 89.
A memory error occurred.
Try to save any readings to the PC.
Then select the
Erase
option from the
Data
menu and then try resetting the meter by
removing battery power for at least 30 seconds.
See also To Cold Boot the HH2 on page 86.
If the problem persists please ask your supplier how to return the unit for repair.
No Sensor
“No Sensor” means that no sensor could be detected when taking a reading. This
could be due to poor cabling.
No Table
“No Table” means table or calibration is missing. When the unit first initialises, or
when a battery is changed and the unit reinitialises, linearisation tables for each
sensor are automatically loaded by the unit. Custom tables can then be loaded by the
user.
If a default table or a custom table subsequently becomes corrupt and you attempt to
take a reading, the error “No Table” is displayed.
Over-range
The sampled voltage value was higher than the HH2 can handle, or the resultant soil
moisture value was calculated to be greater than 1. As soil moisture is a ratio, this is
reported as too high.
Please Change Battery
The battery voltage has dropped below the safety warning threshold of 6.6V.
Change your battery, (but first retrieve your data) :
see Changing the Battery on page 78.
Taking further readings may add a sufficient additional load (most probably with a
PR1) to the battery that will cause the voltage to drop below the CPU operational
threshold. If that happens the integrity of the readings cannot be guaranteed as the
lower voltage might cause the CPU to reset and/or corrupt readings.
The error is a warning only, and is displayed directly after power on and does not
prevent the user using the unit in any way.
The battery is only measured when the unit is woken so to properly check your
battery status allow the HH2 to sleep first, then press ESC to wake it.
Sys Error
See HH2 displays ‘Sys Error’ on page 87.
Summary of Contents for HH2
Page 1: ...User Manual for the Moisture Meter type HH2 Delta T Devices Ltd Version 4 0 1 ...
Page 17: ...10 Navigating Options and Readings Moisture Meter User Manual 4 0 ...
Page 25: ...18 SM200 and ThetaProbes Moisture Meter User Manual 4 0 ...
Page 35: ...28 Profile Probes Moisture Meter User Manual 4 0 ...
Page 45: ...38 Profile Probes Moisture Meter User Manual 4 0 ...
Page 57: ...50 WET Sensor Moisture Meter User Manual 4 0 ...
Page 59: ...52 Equitensiometers Moisture Meter User Manual 4 0 ...