Navigating Options and Readings
15
3. Press the
hash #
key once to display the current cumulative average.
PR2 Avg #?
0.369 100mm
Display Current average
screen
.
#?
indicates that the latest reading is not yet included.
If no average is currently stored then “”No average” is displayed on the bottom line.
4. Press
#
again to update the average with the new reading(s).
PR2 Avg
0.246 100mm
Display Updated Average
Alternatively press
[Esc]
to return to the previous display reading screen without
updating the average.
Once in the
Display Updated Average
screen, press
Esc
to return to the
Display
Reading
screen that you came from. Press
#
to come back to the
Display Updated
Average
screen.
Once the average has been updated, press
Read
while in the
Display Updated
Average
screen to take a new reading.
(If nothing happens press
Esc.
The question
Store?
may appear, reminding you to
decide whether to store the reading. Answer the question by pressing
Store
or
Esc
,
then press
Read
and it should work. See items 6 & 7 below)
What you need to know about averaging
1. Averages are not stored and cannot be uploaded with readings to your PC.
Write them down if you want a permanent record.
2. To erase the current cumulative average press
Esc
until you return to the main
“Delta-T Devices” menu.
3. After each reading, to store the reading use
Store.
To ignore it press
Esc
. (Neither choice affects the average)
4. To add a displayed reading to the cumulative average press the # key twice.
Press
#
once to see
Avg# ?
and the previous cumulative average displayed.
Press
# again,
the question mark disappears and the average is updated.
5. Averaging works whether or not you store individual readings. The process of
selecting readings for inclusion in the cumulative average is completely
independent of the process of storing readings for later upload to your PC
6. You may view the average either before or after deciding to store or erase a
reading.
Summary of Contents for HH2
Page 1: ...User Manual for the Moisture Meter type HH2 Delta T Devices Ltd Version 4 0 1 ...
Page 17: ...10 Navigating Options and Readings Moisture Meter User Manual 4 0 ...
Page 25: ...18 SM200 and ThetaProbes Moisture Meter User Manual 4 0 ...
Page 35: ...28 Profile Probes Moisture Meter User Manual 4 0 ...
Page 45: ...38 Profile Probes Moisture Meter User Manual 4 0 ...
Page 57: ...50 WET Sensor Moisture Meter User Manual 4 0 ...
Page 59: ...52 Equitensiometers Moisture Meter User Manual 4 0 ...