32
Profile Probes
Moisture Meter User Manual 4.0
Soil Set-Up (sub-menu)
Displays available soil parameters that may be defined.
a0 & a1
These options are available for user-defined soils only, not Mineral or Organic
Options: a0 Range: -49.9 to +50.0 Default: 1.0
a1 Range: 0.1 to 50.0 Default: 7.0
Previously you should have selected a custom soil name from the
Options
,
Soil Set-
Up
menu.
Select
a0
. Choose an
a0
value by scrolling through the range and press
Set
.
Select
a1
and repeat the last step.
Field Capacity
Sets the Field Capacity, used for mm Deficit calculations.
Options: 0 to 1 Default: 0.38 Units: m
3
.m
-3
If you want
mm Deficit
you must define the
Field Capacity
(and Root Depth in the
Options, Data
menu).
See the table for Mineral soils on page 106.
Organic soils are so variable you have to determine field capacity experimentally,
before it makes any sense to take mm Deficit readings.
See "Calculating Field Capacity" on page 108.
Select
Options, Soil Set-Up
and choose a soil.
Select
Capacity
.
Scroll to select the actual field capacity.
Press
Set
to select it.
Note : A negative "mm Deficit" result indicates that you set the Field Capacity too low.
Summary of Contents for HH2
Page 1: ...User Manual for the Moisture Meter type HH2 Delta T Devices Ltd Version 4 0 1 ...
Page 17: ...10 Navigating Options and Readings Moisture Meter User Manual 4 0 ...
Page 25: ...18 SM200 and ThetaProbes Moisture Meter User Manual 4 0 ...
Page 35: ...28 Profile Probes Moisture Meter User Manual 4 0 ...
Page 45: ...38 Profile Probes Moisture Meter User Manual 4 0 ...
Page 57: ...50 WET Sensor Moisture Meter User Manual 4 0 ...
Page 59: ...52 Equitensiometers Moisture Meter User Manual 4 0 ...