84
Technical Reference
Moisture Meter User Manual 4.0
25-way D-connector
The Moisture Meter uses a 25-way male connector to connect to either a PC (via an
RS232 cable) or to a sensor.
The following diagram indicates which of the pins on the D-connector are used for
which function. The HH2 has been designed to avoid sharing pins for the sensor and
PC connections to prevent accidental damage to attached equipment.
See earlier sections of this document for a detailed description of the connector
accessories, which Delta-T can provide for connecting sensors to the HH2.
25-way D-connector Pin Arrangement
Note: the pins shown here are those seen from viewing the D-connector on the HH2
when holding the meter in your hand.
All other pins are unused. They are not connected within the HH2.
See also 25-pin D Socket Kit on page 99.
VBAT Power out +
13
25
AN1
Sensor 1 Voltage in (+'ve)
XIO 0 unused digital I/O
12
24
AN2
Sensor 2 Voltage in (+'ve)
XIO 1 unused digital I/O
11
23
AN3
Sensor 3 Voltage in (+'ve)
Unused
10
22
Unused
Unused
9
21
AN4
Sensor 4 Voltage in (+'ve)
Unused
8
20
Unused
Ground
7
19
Unused
DSR
6
18
AN5
Sensor 5 Voltage in (+'ve)
CTS
5
17
Unused
RTS
4
16
AN6
Sensor 6 Voltage in (+'ve)
RxD
3
15
Unused
TxD
2
14
Analogue Ground
Ground
1
Summary of Contents for HH2
Page 1: ...User Manual for the Moisture Meter type HH2 Delta T Devices Ltd Version 4 0 1 ...
Page 17: ...10 Navigating Options and Readings Moisture Meter User Manual 4 0 ...
Page 25: ...18 SM200 and ThetaProbes Moisture Meter User Manual 4 0 ...
Page 35: ...28 Profile Probes Moisture Meter User Manual 4 0 ...
Page 45: ...38 Profile Probes Moisture Meter User Manual 4 0 ...
Page 57: ...50 WET Sensor Moisture Meter User Manual 4 0 ...
Page 59: ...52 Equitensiometers Moisture Meter User Manual 4 0 ...