Programmers
Replacing a Programmer
PSV5000 Owner’s Manual
- 121 -
7.
On the Diagnostics window, select the programmer with the DAB
installed.
8.
Click Test All.
The pass/fail test results are displayed in TaskLink and are also written
to /fdrroot/system/diaglog.txt (and to the CF-card of the target
FlashCORE programmer). These files can be viewed in TaskLink and
saved (on the Handler Computer).
If any of the tests show Fail in the TaskLink display, contact your nearest
Data I/O Service Center for repair options. To help our service personnel
diagnose the problem, please e-mail both the
eventlog.txt and diaglog.txt files.
Sample
dialog.txt
file:
Diagnostic Pass #1
Run Vpp Overcurrent test.
DUT 1 G1 Vpp overcurrent is sensed at 57 mA.
DUT 1 G2 Vpp overcurrent is sensed at 57 mA.
DUT 1 G3 Vpp overcurrent is sensed at 57 mA.
DUT 1 G4 Vpp overcurrent is sensed at 57 mA.
DUT 2 G1 Vpp overcurrent is sensed at 57 mA.
DUT 2 G2 Vpp overcurrent is sensed at 57 mA.
DUT 2 G3 Vpp overcurrent is sensed at 57 mA.
DUT 2 G4 Vpp overcurrent is sensed at 57 mA.
Error: DUT 3 G1 Vpp overcurrent is not sensed from 50 mA to 70
mA.
Error: DUT 3 G2 Vpp overcurrent is not sensed from 50 mA to 70
mA.
Error: DUT 3 G3 Vpp overcurrent is not sensed from 50 mA to 70
mA.
Error: DUT 3 G4 Vpp overcurrent is not sensed from 50 mA to 70
mA.
DUT 4 G1 Vpp overcurrent is sensed at 57 mA.
DUT 4 G2 Vpp overcurrent is sensed at 57 mA.
DUT 4 G3 Vpp overcurrent is sensed at 57 mA.
DUT 4 G4 Vpp overcurrent is sensed at 57 mA.
Diagnostics failed.
Summary of Contents for PSV5000
Page 12: ......
Page 36: ......
Page 160: ......
Page 161: ...Electrical Schematics for PSV5000 and Equipment ...
Page 162: ......
Page 163: ......
Page 164: ......
Page 165: ......
Page 166: ......
Page 167: ......
Page 168: ......
Page 169: ......
Page 170: ......
Page 171: ......
Page 172: ......
Page 173: ......
Page 174: ......
Page 175: ......
Page 176: ...PROGRAMMABLE MEDIA EXPERTS ...