Maintenance ■ Programmers
- 120 -
Data I/O ■ 096-0465-001C
The wrist strap Data I/O
PN is 440-0021-001.
R
UNNING THE
P
ROGRAMMER
D
IAGNOSTIC
T
EST
To run diagnostic tests on FlashCORE programmer(s):
1.
Finish a job if one is running; wait for the PNP head to park.
2.
Turn OFF the programmer circuit breaker at the Power Panel (down
position).
3.
Exit CH700 to return to TaskLink.
WARNING: ESD hazard! To prevent ESD shock, before you
touch the Socket Adapter, discharge static electricity from
yourself by touching a common ground or an unpainted metal
surface.
When at the PSV5000 Machine, always wear a wrist strap containing a 1
M-ohm min. to 10 M-ohm max. current limiting resistor. Connect the
antistatic wrist strap to the grounding socket on the front or back of the
PSV5000 Machine.
WARNING: Possible collision hazard! The high speed and force of the gantry
can seriously harm anyone working inside the workspace.
When working within the machine workspace, moving the PNP head must
be the responsibility of only one qualified individual. All others must stay
clear of the machine controls to prevent injury to that person.
Ensure that a job is Paused or Finished, or the system power is OFF prior to
opening any safety doors.
4.
Remove the Socket Adapter from the target programmer and insert the
Diagnostic Adapter Board, ensuring that it aligns correctly on the adapter
pins. Screw down the two bracket screws (4 mm hex key).
5.
Turn the programmer circuit breaker at the Power Panel ON (up
position).
6.
Start TaskLink and click Tools > Run Programmer Diagnostics.
Summary of Contents for PSV5000
Page 12: ......
Page 36: ......
Page 160: ......
Page 161: ...Electrical Schematics for PSV5000 and Equipment ...
Page 162: ......
Page 163: ......
Page 164: ......
Page 165: ......
Page 166: ......
Page 167: ......
Page 168: ......
Page 169: ......
Page 170: ......
Page 171: ......
Page 172: ......
Page 173: ......
Page 174: ......
Page 175: ......
Page 176: ...PROGRAMMABLE MEDIA EXPERTS ...