Performing a Manual Probe Zero (On Block)
Note:
When the zero probe option is set to manual, the probe zero disk
(battery cap) located on the top of the gauge, will be used as a zero standard
and the warning screen illustrated above will be displayed.
1) Press the
OK
or
ESC
keys to enter the main measurement screen and
begin the manual zero process.
2) Apply a drop of couplant on the transducer and place the transducer in
steady contact with the probe zero disk, and obtain a steady reading.
3) Press the
MENU
key once to activate the menu items tab. Press the
MENU
key multiple times to tab right and the
ESC
key multiple times to tab left until
the
PRB
menu is highlighted and displaying the submenu items.
4) Press the
UP
and
DOWN
arrow keys to scroll through the sub menu items
until
ZERO PROBE
is highlighted.