DFX-7
Ultrasonic Flaw Detector
11
Coating Probe Identified
Coating Probe Not Identified
5) Press the
ENTER
key to display the confirmation screen.
6) If a coating transducer was identified use the
UP
and
DOWN
arrow keys to
toggle coating on/off.
7) Press the
OK
key to complete the probe zero function, or
ESC
key to cancel
the probe zero function.
8) Remove the transducer from the probe zero disk, and proceed to the
calibration section.
Note:
The value that is displayed will change depending on the current velocity
setting in the
DFX-7
. Disregard the number that is displayed.
It is not
important
. What is important is accurately performing the steps outlined above
to insure reliability of the probe zero calculation.
One Point Material Calibration
For the purposes of this quick start section, we’ll only be covering the most common
one point calibration option to determine the sound velocity of the test material. It
would be very handy to carry a set of mechanical calipers to use in conjunction with
the
DFX-7
for calibration in the field: