DFX-7
Ultrasonic Flaw Detector
71
5) Press the
LEFT
and
RIGHT
arrow keys to scroll the digit locations.
6) Repeat steps
4 & 5
until the
DELAY
value is correctly displayed.
7) Press the
OK
key to return to the measurement screen, or
ESC
to cancel
entering the
DELAY
.
8) Finally, press the
MEAS
key to return to the measurement screen and begin
taking readings.
Adjusting the Range (B-Depth)
Once again, just to reiterate, the
Range (B-Depth) + Delay (B-Start) equals the
right side of the screen
. Therefore, the Range (B-Depth) is the overall area, from
the delay, that will be viewable on the screen. The
DFX-7
digitizer will round up from
the Range (B-Depth) that’s entered. Therefore, if the Range (B-Depth) is set at 1.0”,
the digitizer will round this value up to the next adjustment available. The procedures
to adjust the overall thickness range viewed Range (B-Depth) are outlined below:
Adjusting the Range (B-Depth) using the Hot Menus
1) Press the
MEAS
key once to activate measure menu items. Press the
MEAS
key multiple times to move right and the
ESC
key multiple times to
move left, until the
WIDTH
cell is highlighted.