Dakota Ultrasonics
122
point (C)
is the peak hold symbol - carrot. The carrot is only displayed if the flaw
mode is set to “peak”. The peak hold position represents the highest amplitude
obtained during the scan. Pressing the
CLR
key will reset the peak hold position to
zero.
Enabling the Flaw Mode:
Note:
This mode requires an accessory dual lemo to single microdot cable is
available from the factory. The procedures to use the
FLAW MODE
feature are
outlined below:
Using the Default Setup
1) Attach the special dual lemo to microdot cable and angle beam transducer
to the
DFX-7
.
2) Press the
MENU
key once to activate the menu items tab. Press the
MENU
key multiple times to tab right and the
ESC
key multiple times to tab left until
the
PROBE
menu is highlighted and displaying the submenu items.
3) Use the
UP
and
DOWN
arrow keys to scroll through the sub menu items
until
FLAW MODE
is highlighted.
4) Use the
LEFT
and
RIGHT
arrow keys to scroll the flaw mode options (off,
on, peak).