Using a Known Thickness
Note:
Be sure that the probe zero procedure has been performed prior to
performing this calibration procedure.
1) Physically measure an exact sample of the material or a location directly on
the material to be measured using a set of calipers or a digital micrometer.
2) Apply a drop of couplant on the transducer and place the transducer in
steady contact with the sample or actual test material. Be sure that the
reading is stable and the repeatability indicator, in the top left corner of the
display, is fully lit and stable. Press the
MENU
key once to activate the
menu items tab. Press the
MENU
key multiple times to tab right and the
ESC
key multiple times to tab left until the
CAL
menu is highlighted and
displaying the submenu items.
3) Use the
UP
and
DOWN
arrow keys to scroll through the sub menu items
until
MATL 1PT
is highlighted.