Dakota Ultrasonics
106
3) Use the
UP
and
DOWN
arrow keys to scroll through the sub menu items
until
COATING 1PT
is highlighted.
4) Press the
ENTER
key to display the
Digits Edit Box
.
5) Press the
UP
and
DOWN
arrow keys to scroll the highlighted value.
6) Press the
LEFT
and
RIGHT
arrow keys to scroll the digit locations.
7) Repeat steps
5 & 6
until the known thickness value is correctly displayed.
8) Press the
OK
key to calculate the velocity and return to the menu screen, or
ESC
to cancel the coating one point calibration.
9) Finally, press the
MEAS
key to return to the measurement screen and begin
taking readings.