Dakota Ultrasonics
16
quick start section only the 1pt option
PECT (pulse-echo coating)
mode will be
covered. Refer to the calibration section of the manual for a complete explanation on
the coating calibration options. The following steps below outline the necessary
steps to either set the velocity of the coating, or perform a one point calibration to
calculate the coating velocity:
Known Coating Velocity
If the coating velocity is known, the user may wish to simply enter the velocity
number into the
DFX-7
, rather than have the
DFX-7
calculate the velocity value using
a known thickness on a coating sample. The steps for entering the velocity are
outlined below:
Using a Known Coating Velocity
1) Press the
MENU
key once to activate the menu items tab. Press the
MENU
key multiple times to tab right and the
ESC
key multiple times to tab left until
the
CAL
menu is highlighted and displaying the submenu items.
2) Use the
UP
and
DOWN
arrow keys to scroll through the sub menu items
until
COATING VEL
is highlighted.
3) Press the
ENTER
key to display the
Digits Edit Box
.
4) Press the
UP
and
DOWN
arrow keys to scroll the highlighted value.
5) Press the
LEFT
and
RIGHT
arrow keys to scroll the digit locations.