DFX-7
Ultrasonic Flaw Detector
75
6) Finally, press the
MEAS
key to return to the measurement screen and begin
taking readings.
10.4 Auto Find
The
DFX-7
is equipped with a very handy feature for locating the detection point of a
return echo that is not currently in the viewable region of the display. The
AUTO
FIND
feature is a convenient way to automatically bring the detected echo into view
on the display. Once the echo is in the viewable area, the user can then make small
adjustments to the delay and range accordingly. The procedure to use the
AUTO
FIND
feature is outlined below:
Using Auto Find
1) Press the
MENU
key once to activate the menu items tab. Press the
MENU
key multiple times to tab right, and the
ESC
key multiple times to tab left,
until the
UTIL
menu is highlighted and displaying the submenu items.
2) Use the
UP
and
DOWN
arrow keys to scroll through the sub menu items
until
AUTO FIND
is highlighted.
3) Place the transducer on the test material and obtain a reading.