2 .Checking and adjusting the circular level
●
Check
Carefully level the instrument with the plate level. If the bubble of the circular level is
centered properly at this time, adjustment is not required. Otherwise, proceed with the
following adjustment.
●
Adjustment
Shift the bubble to the center of the level by adjusting three capstan adjustment screws
on the bottom surface of the circular level, with the accessory adjusting pin.(see
diagram)
Adjustment screws
24.2 Reticle
Bottom of the base
●
Check
①
Set the instrument on the tripod and carefully level it.
②
Sight the cross-hair on a well-defined point A on the wall at a distance of at least
50 meters. (160ft )
③
Next swing the telescope and check whether the point travels along the length of
the vertical cross-hair.
④
If the point appears to move continuously on the vertical hair, the vertical cross-hair
lies in a plane perpendicular to the horizontal axis.(adjustment is not required.)
●
Adjustment
①
Unscrew the cross-hair adjustment section cover by revolving it in the counterclockwise
direction, and take it off. This will expose four eyepiece section attachment screws.
②
Loosen all four attachment screws slightly with the accessory screw-driver. (while
taking note of the number of the revolutions) make vertical cross-hair coincide with A by
turning eyepiece and tighten the four attachment screws.
102
Attachment screws
Eye Piece
Attachment screws