24. CHECKS AND ADJUSTMENTS
The total station is a precision instrument that requires fine adjustments. It must be
inspected and adjusted before use so that it always performs accurate measurements.
Using “24.1 Plate level and circular level”
,
Check to ensure that the bubbles are accurately
in adjustment.
In addition, the instrument should be inspected with special care after it has been stored a
long time, transported, or when it may have been damaged by a strong shock.
24.1 Plate level and circular level
1
.
Checking and adjusting the plate level
●
Check
①
Place the plate level parallel to a line running through the centers of two
leveling screws (eg. A, B). Use these two screws to place the bubble in the
center of the plate level vial.
②
Rotate the instrument 180°or 200g around the vertical axis and check bubble
movement of the plate level. If the bubble has been displaced, then proceed with
the following adjustment
.
●
Adjustment
①
Adjust the level adjustment capstan screw , with the accessory adjusting pin and
return the bubble towards the center of the plate level vial. However, correct only
one-half of the displacement by this method.
②
Correct the remaining 1/2 amount of the bubble displacement with the leveling
screws.
③
Rotate the instrument 180°or 200g around the vertical axis once more and check bubble
movement. if the bubble is still displaced, then repeat the adjustment.
101
Half amount of displacement
Level adjustment screw