Continuum Surelite Laser
50
996-0207
Doubler Type 2
619-2640
619-2640
619-2640
10
Tripler
619-2720
619-2720
619-2720
11
Dichroics 532 nm
Dichroics 355 nm
Dichroics 266 nm
105-0022
105-0023
105-0025
105-0022
105-0023
105-0025
105-0022
105-0023
105-0025
12
Dichroics 1,064 nm
105-0002
105-0002
105-0002
13
Beam dump
619-1437
619-1437
619-1437
Laser heads
General description
Continuum’s Surelite lasers use the 811U-06 head for the Surelite I, the 811U-07 head for the
Surelite II models, and a 821V-09 head for the Surelite III model. They all have a beam height of
43 mm. The head is a modular design incorporating a rod, a flashlamp and coupling medium. It
has a linear flashlamp which is in a close coupled configuration surrounded by a high brilliance
magnesium oxide diffuser. This results in a high pumping efficiency that minimizes thermal loading
and reduces power consumption. These features combine to create the excellent pumping
homogeneity required for producing high gain, superior quality beams.
Rods
The head is designed to pump Nd+3 doped YAG. The doping levels vary from 0.9 to 1.4%. The
YAG rod ends have hard dielectric antireflective coatings (AR). The rod length is 115 mm
measured along the optical axis. The diameter of the 811U-06 rod is 6 mm and the 812V-09 rod is
9 mm.
Flashlamps
The heads are pumped by linear flashlamps. The lamp has a voltage polarity that must be observed.
The flashlamp anode can be identified by a red marking on the lamp electrode whereas the cathode
has no marking. The discharge system of the flashlamp uses a negative, critically damped pulse (-1.8
kV max) with a duration of 200 microseconds full width half max (FWHM). The gas in the
flashlamp is xenon with a pressure of 1-3 atmospheres.
Cooling
Flashlamps and laser rods must be cooled by distilled, deionized water. Resistivity of distilled,
deionized cooling water should not fall below 500 k
Ω
/cm. Do not mix any chemical products with
the distilled water.
Summary of Contents for Surelite EX 10
Page 8: ...Continuum Surelite Laser 6 996 0207 This page is intentionally blank...
Page 12: ...Continuum Surelite Laser 2 996 0207 This page is intentionally blank...
Page 26: ...Continuum Surelite Laser 16 996 0207...
Page 36: ...Continuum Surelite Laser 26 996 0207 This page is intentionally blank...
Page 57: ...Section 3 System Operation 996 0207 47 Figure 27 Surelite chronograph...
Page 80: ...Continuum Surelite Laser 70 996 0207 This page is intentionally blank...
Page 92: ...Continuum Surelite Laser 82 996 0207...
Page 94: ...Continuum Surelite Laser 84 996 0207 This page is intentionally blank...
Page 106: ...Continuum Surelite Laser 96 996 0207 This page is intentionally blank...
Page 107: ...Appendix B Operator Notes...
Page 108: ...Continuum Surelite Laser 98 996 0207...
Page 109: ...Appendix B Operator Notes 996 0207 99...
Page 110: ...Continuum Surelite Laser 100 996 0207...
Page 111: ...Appendix B Operator Notes 996 0207 101...
Page 112: ...Continuum Surelite Laser 102 996 0207 This page is intentionally blank...