Section 6: Troubleshooting
996-0207
79
Figure 53 Flashlamp rise – free-
running pulse
Figure 54 Flashlamp rise – further
adjustment required
Note:
To eliminate the free-running, adjustments to the polarization optics
need to be made.
11. At this time make a small adjustment to the
vertical axis of the Pockels cell through the
vertical adjustment access hole located at the
top of the Pockels cell mount.
12. Note the vertical amplitude of the scope
display. If it becomes less, then the previous
adjustment was made in the correct direction.
13.
Continue to adjust the vertical axis until the amplitude of the trace is as small as
possible or until the free-running stops.
If
the free-running stops, then go on to step
16.
14. Now follow the same procedure as in steps 12
and 13, only this time adjust the horizontal
axis located on the lower corner of the
Pockels cell mount.
15. Continue these adjustments until the free-running has been eliminated.
16. Close the intracavity shutter and increase the pump voltage by 100 V.
CAUTION!
Do not exceed 1.8 kV unless otherwise instructed
to do so by a Continuum representative.
17. Open the intracavity shutter and look at the oscilloscope to see if the laser is free-
running.
18. If the laser is in a free-running state, then repeat steps 11 through 15 while at this
higher pump voltage.
Note
: Once the oscillator optics are properly adjusted, passive free-running should
not occur until the pump voltage has exceeded 100 V above the normal operating
level. In many cases this voltage “ buffer” will be even greater than 100 V.
19. Once the “ buffer” voltage is ~100 V, return the pump voltage to its normal level. (See
QA chart in front of the manual immediately following the title page.)
Note
: After completing the free-running test procedure, it is very important to
check the oscillator alignment once again. Any movement of the Pockels cell may
cause a misalignment, requiring a back mirror adjustment (see page 6-6).
Flashlamp voltage adjustment
It is common for flashlamps to last 20-30 million shots, but during that time the efficiency of
the lamps will decline. Initially, it is possible to compensate for the decline by increasing
pump voltage. However, after a point, the thermal loading on the YAG rod will adversely
Summary of Contents for Surelite EX 10
Page 8: ...Continuum Surelite Laser 6 996 0207 This page is intentionally blank...
Page 12: ...Continuum Surelite Laser 2 996 0207 This page is intentionally blank...
Page 26: ...Continuum Surelite Laser 16 996 0207...
Page 36: ...Continuum Surelite Laser 26 996 0207 This page is intentionally blank...
Page 57: ...Section 3 System Operation 996 0207 47 Figure 27 Surelite chronograph...
Page 80: ...Continuum Surelite Laser 70 996 0207 This page is intentionally blank...
Page 92: ...Continuum Surelite Laser 82 996 0207...
Page 94: ...Continuum Surelite Laser 84 996 0207 This page is intentionally blank...
Page 106: ...Continuum Surelite Laser 96 996 0207 This page is intentionally blank...
Page 107: ...Appendix B Operator Notes...
Page 108: ...Continuum Surelite Laser 98 996 0207...
Page 109: ...Appendix B Operator Notes 996 0207 99...
Page 110: ...Continuum Surelite Laser 100 996 0207...
Page 111: ...Appendix B Operator Notes 996 0207 101...
Page 112: ...Continuum Surelite Laser 102 996 0207 This page is intentionally blank...