996-0207
E08
Lamp simmer not detected.
E09
Flow switch stuck on.
CSE
This is displayed if a “ check sum error” occurs in the EEPROM. This
display means that the EEPROM has lost its saved memory and has reverted
to preset software limits. It will be necessary to toggle through the display
menu and re-enter the QA values for frequency (Fxx), lamp pump voltage, Q-
switch delay and pulse division (P01). If display does not go away after
recycling keyswitch, contact Continuum.
Single shot (SS) button
An operational feature of your Surelite is the ability to operate the system in a single shot
mode. That is, a single laser pulse on demand. This feature can be very useful when aligning
downstream optics or verifying target position
.
To enable SS
To enable the Single Shot operation do the following:
1. Perform the Daily Shutdown procedure found on page
28
.
2. Locate the Single Shot (SS) cable that was shipped with your system in the
Accessories Box. It is approximately 15 ft. (4.5 m) long, has a momentary push
button switch at one end, and a BNC connector at the other.
3. Connect the Single Shot cable to BNC connector labeled
SINGLE SHOT
on front
panel of PCU.
4. Turn the keyswitch to the
ON
position.
5. Press the
SELECT
button and toggle to pulse division (Pxx). Check that the pulse
division displayed is >00. The normal reading for a system would be P01.
Note:
If system operation mode was in P00 prior to initialization and user
wants a mode other than single shot, he must press SELECT ARROW UP
button to obtain desired laser output.
6. Press the
START/STOP
button. Laser head should now flash.
7. Press the
SHUTTER
button on. System will now be lasing.
8. Initially press the SS button once and lasing output will stop but the laser head will
continue to flash.
9. From this point on, each subsequent press of the SS button results in a single laser
pulse.
To disable Single Shot (SS) operation
1. With the system on, the laser head flashing and
SHUTTER
button
ON
, depress
SS
button and hold down for 10 seconds. Normal lasing operation will resume.
Summary of Contents for Surelite EX 10
Page 8: ...Continuum Surelite Laser 6 996 0207 This page is intentionally blank...
Page 12: ...Continuum Surelite Laser 2 996 0207 This page is intentionally blank...
Page 26: ...Continuum Surelite Laser 16 996 0207...
Page 36: ...Continuum Surelite Laser 26 996 0207 This page is intentionally blank...
Page 57: ...Section 3 System Operation 996 0207 47 Figure 27 Surelite chronograph...
Page 80: ...Continuum Surelite Laser 70 996 0207 This page is intentionally blank...
Page 92: ...Continuum Surelite Laser 82 996 0207...
Page 94: ...Continuum Surelite Laser 84 996 0207 This page is intentionally blank...
Page 106: ...Continuum Surelite Laser 96 996 0207 This page is intentionally blank...
Page 107: ...Appendix B Operator Notes...
Page 108: ...Continuum Surelite Laser 98 996 0207...
Page 109: ...Appendix B Operator Notes 996 0207 99...
Page 110: ...Continuum Surelite Laser 100 996 0207...
Page 111: ...Appendix B Operator Notes 996 0207 101...
Page 112: ...Continuum Surelite Laser 102 996 0207 This page is intentionally blank...