Section 4: Subassembly and Module Descriptions
996-0207
59
LASER ON LED.
This red LED comes on when water flow is established.
KEY SWITCH.
Turning the key from off to
ON
position resets system logic and enables
power-up sequence to begin.
START/STOP.
This button switch starts and stops the laser head flashlamp.
SHUTTER BUTTON.
Depressing this switch to the
OPEN
position will open the intra-
cavity shutter on the laser bench and start lasing. Depressing this switch again will
CLOSE
the shutter. An audible click is heard when opening or closing the intra-cavity shutter.
DISPLAY READOUT LED.
This digital display read-out shows all the error messages
and commands. The messages are explained on page
29
. The red
VOLTAGE KV
LED
and
Q -SW DELAY
µs
LED light up when the
DISPLAY
LED shows their readings.
SELECT BUTTON
.
This button allows the user to step through the different operating
parameters which are explained on page
28
. The select
ARROW UP
and select
ARROW
DOWN
buttons are used to change the value of display parameters.
RS232 9-pin male plug.
See page
40
for serial plug for RS232 operation.
EXTERNAL 9 pin female plug.
The optional externals plug for a remote control device:
External interlock loop.
See page
60
.
DAT
(direct access triggering). See page
43
.
SINGLE SHOT.
This connector accepts the Single Shot Cable (SS). When the SS is
plugged in you get one laser pulse each time one manually presses the SS fire button. When
the SS is disconnected the laser will fire at the last saved rep rate.
VARIABLE SYNC OUT.
This connector provides a variable trigger output which is
adjustable via a pot to the right of
VARIABLE SYNC OUT
BNC on front panel of the
PCU. The signal provided is negative going TTL signal (3.5 V to 0 V into a 1
Μ
termination) that can be adjusted from 325 ns before lasing to 450 ns after lasing.
FIXED SYNC OUT.
This BNC connector provides a negative going TTL signal (3.5 V to
0 V into a 1
Μ
termination) approximately 100 ns before laser pulse output.
FLASHLAMP SYNC.
This BNC style connector provides a +3.5 V pulse at the start of
the firing of the flashlamp. This signal occurs ~180 µs before lasing output (depends on
Q-
SWITCH DELAY
setting).
Rear panel
DI IN.
The distilled water input hose connector carrying warm water from laser head back
to cooling group.
DI OUT.
The distilled water output hose connector carrying water from cooling group to
laser head.
Summary of Contents for Surelite EX 10
Page 8: ...Continuum Surelite Laser 6 996 0207 This page is intentionally blank...
Page 12: ...Continuum Surelite Laser 2 996 0207 This page is intentionally blank...
Page 26: ...Continuum Surelite Laser 16 996 0207...
Page 36: ...Continuum Surelite Laser 26 996 0207 This page is intentionally blank...
Page 57: ...Section 3 System Operation 996 0207 47 Figure 27 Surelite chronograph...
Page 80: ...Continuum Surelite Laser 70 996 0207 This page is intentionally blank...
Page 92: ...Continuum Surelite Laser 82 996 0207...
Page 94: ...Continuum Surelite Laser 84 996 0207 This page is intentionally blank...
Page 106: ...Continuum Surelite Laser 96 996 0207 This page is intentionally blank...
Page 107: ...Appendix B Operator Notes...
Page 108: ...Continuum Surelite Laser 98 996 0207...
Page 109: ...Appendix B Operator Notes 996 0207 99...
Page 110: ...Continuum Surelite Laser 100 996 0207...
Page 111: ...Appendix B Operator Notes 996 0207 101...
Page 112: ...Continuum Surelite Laser 102 996 0207 This page is intentionally blank...