CAUTION!
Never attempt to adjust the output coupler
(the mirror mount nearest to the exit port).
11. Once normal power has been restored, shut down the system, remove the interlock
defeat bracket and reinstall oscillator housing cover.
12. Resume normal operation.
Oscillator optimization (with an oscilloscope)
For the Surelite series lasers that employ the “ Gaussian” oscillator, optimizing cavity
alignment using an oscilloscope can be very useful.
If the oscillator is well aligned the output pulsewidth is at its narrowest (5-8 ns). The converse
is also true. When the cavity is misaligned the pulsewidth will broaden. Therefore,
monitoring the pulse width while making oscillator adjustments will allow you to achieve
optimum cavity alignment.
N
ECESSARY
E
QUIPMENT
•
Burn paper
•
Clear plastic bags
•
High density Styrofoam beam dump
•
Photodiode with ~1 ns rise time
•
BNC cable
•
>350 MHz oscilloscope with enhanced view screen
This procedure optimizes the oscillator by monitoring the pulsewidth.
Note:
This procedure should be done in conjunction with the alignment
procedure described previously in this section.
1. Remove harmonic crystals.
2. Place the beam block in the beam path after the output port.
3. Set up the oscilloscope:
•
Plug in photodiode to 50
Ω
termination.
•
Run a trigger BNC cable from the “ Fixed sync out” BNC (located on the PCU
front panel display) to the time base of the scope.
•
Set the triggering of the scope to trigger off the +5 V TTL sync pulse.
•
Adjust the time base to view a 5-10 ns pulse. The trigger pulse will occur ~90 ns
before lasing.
•
Place the photodiode on the bench so that it will be able pick up the laser light
scattered off the beam block.
Be sure that the voltage setting on the power unit is set to the laser’s normal operating
voltage.
Summary of Contents for Surelite EX 10
Page 8: ...Continuum Surelite Laser 6 996 0207 This page is intentionally blank...
Page 12: ...Continuum Surelite Laser 2 996 0207 This page is intentionally blank...
Page 26: ...Continuum Surelite Laser 16 996 0207...
Page 36: ...Continuum Surelite Laser 26 996 0207 This page is intentionally blank...
Page 57: ...Section 3 System Operation 996 0207 47 Figure 27 Surelite chronograph...
Page 80: ...Continuum Surelite Laser 70 996 0207 This page is intentionally blank...
Page 92: ...Continuum Surelite Laser 82 996 0207...
Page 94: ...Continuum Surelite Laser 84 996 0207 This page is intentionally blank...
Page 106: ...Continuum Surelite Laser 96 996 0207 This page is intentionally blank...
Page 107: ...Appendix B Operator Notes...
Page 108: ...Continuum Surelite Laser 98 996 0207...
Page 109: ...Appendix B Operator Notes 996 0207 99...
Page 110: ...Continuum Surelite Laser 100 996 0207...
Page 111: ...Appendix B Operator Notes 996 0207 101...
Page 112: ...Continuum Surelite Laser 102 996 0207 This page is intentionally blank...