Section 3: System Operation
Daily Startup
1. Check the water reservoir level before beginning
Startup
for the first time.
2.
The toggle switch labeled,
AC Power
, should have been left on after the last use. If
not, do the following:
•
On the front panel of the Surelite power unit, flip the toggle switch up that is
labeled
AC P
OWER
(this is the
On
position). After a quick diagnostic check, the
LEDs on the front panel display will readout 888, software rev. #, then
O
FF
.
The “Power On” button is located on the rear panel of the laser head.
•
After approximately 6 seconds, the
L
ASER
O
N
lights will turn on. They are
located on the top of the laser head and on the front panel of the power supply.
3. Insert the key into the keyswitch and rotate it counter-clockwise to the On postion.
The readout LED will monmentarily display the following.
Total # flashlamp discharges
Displays as a series of 3 displays for millions,
thousands, and hundreds.
Water flow startup
FLo displays while system waits for water flow
to be verified in the laser heads.
Flashlamp discharge
frequency
Displays the last saved rep, in Hertz
(flashes/second).
4. Press the
SELECT
button and toggle the LED display through these six modes to
verify the accuracy of all the settings. The normal values are:
Rep Rate
Factory set for either 10 or 20 Hz
Q-switch
Set to the value listed in the Quality Assurance
Document at the front of the manual
Pulse Division
P01 (laser output with every discharge of
flashlamp)
Pump Voltage
Set to the value listed in the Quality Assurance
Document at front of manual.
Serial Mode
SoF (serial port off)
External Mode
EoF (External mode off)
5. Press the
START/STOP
button and its LED will come on. The laser head will now
flash at the designated frequency. Allow the system to run for 15-20 minutes to
thermally stabilize the YAG head before proceeding.
6. Put on laser safety goggles.
7. On the front of the laser bench, slide the exit port shutter to the
OPEN
position.
LASER RADIATION!
The next step results in laser output.
8. Press the
SHUTTER
button and its LED will come on. An audible click will be
heard from laser bench as the intra-cavity shutter withdraws from the oscillator. The
system should now lase.
Summary of Contents for Surelite EX 10
Page 8: ...Continuum Surelite Laser 6 996 0207 This page is intentionally blank...
Page 12: ...Continuum Surelite Laser 2 996 0207 This page is intentionally blank...
Page 26: ...Continuum Surelite Laser 16 996 0207...
Page 36: ...Continuum Surelite Laser 26 996 0207 This page is intentionally blank...
Page 57: ...Section 3 System Operation 996 0207 47 Figure 27 Surelite chronograph...
Page 80: ...Continuum Surelite Laser 70 996 0207 This page is intentionally blank...
Page 92: ...Continuum Surelite Laser 82 996 0207...
Page 94: ...Continuum Surelite Laser 84 996 0207 This page is intentionally blank...
Page 106: ...Continuum Surelite Laser 96 996 0207 This page is intentionally blank...
Page 107: ...Appendix B Operator Notes...
Page 108: ...Continuum Surelite Laser 98 996 0207...
Page 109: ...Appendix B Operator Notes 996 0207 99...
Page 110: ...Continuum Surelite Laser 100 996 0207...
Page 111: ...Appendix B Operator Notes 996 0207 101...
Page 112: ...Continuum Surelite Laser 102 996 0207 This page is intentionally blank...