Continuum Surelite Laser
80
996-0207
affect beam quality. This procedure instructs you on how to increase flashlamp pump voltage
to compensate for lamp aging.
N
ECESSARY
E
QUIPMENT
•
power meter
•
IR card
S
TEPS
1. Before increasing the pump voltage to bring the laser output back to spec energy, make
sure that your power decrease is not due to misalignment or optical damage.
2. Verify the energy specs by reviewing the Error! Reference source not found. on page
Error! Bookmark not defined. the front of the manual.
3. Position a power meter to monitor the IR energy. If the system has optional
harmonics, measure the total energy of the combined wavelengths. Due to losses in
the harmonic crystals, expect the measured combined total to be 15-20% below
published spec for IR alone.
4. Increase the pump voltage by 20 V on the PCU by the pressing the
SELECT
button
until red LED is lit next to
VOLT-AGE KV
on the front panel of the PCU.
5. Now use the
ARROW UP/ARROW DOWN
buttons as needed to reach +20 V.
6. Open the intra-cavity shutter. Allow the laser to Q-switch and measure the energy.
7. If still below specs, repeat steps 4 thru 6.
Note
:
If the total voltage increases exceed the installed voltage by more than
100 V, then the flashlamps should be replaced.
8. Check the free-running buffer. It needs to be at least 100 V. See
Eliminating free-
running
on page
77
.
9. If the energy is back to its spec level, log the date and action in “ Operator Notes” at
the end of the manual and the procedure is complete.
Cooling group contamination
This section describes the procedure for flushing the cooling group if it becomes
contaminated with organic material (algae).
N
ECESSARY
E
QUIPMENT
•
1 pint (~500 ml) 30% hydrogen peroxide
•
3 gal. distilled water
•
1 DI filter (PN#313-0099)
Summary of Contents for Surelite EX 10
Page 8: ...Continuum Surelite Laser 6 996 0207 This page is intentionally blank...
Page 12: ...Continuum Surelite Laser 2 996 0207 This page is intentionally blank...
Page 26: ...Continuum Surelite Laser 16 996 0207...
Page 36: ...Continuum Surelite Laser 26 996 0207 This page is intentionally blank...
Page 57: ...Section 3 System Operation 996 0207 47 Figure 27 Surelite chronograph...
Page 80: ...Continuum Surelite Laser 70 996 0207 This page is intentionally blank...
Page 92: ...Continuum Surelite Laser 82 996 0207...
Page 94: ...Continuum Surelite Laser 84 996 0207 This page is intentionally blank...
Page 106: ...Continuum Surelite Laser 96 996 0207 This page is intentionally blank...
Page 107: ...Appendix B Operator Notes...
Page 108: ...Continuum Surelite Laser 98 996 0207...
Page 109: ...Appendix B Operator Notes 996 0207 99...
Page 110: ...Continuum Surelite Laser 100 996 0207...
Page 111: ...Appendix B Operator Notes 996 0207 101...
Page 112: ...Continuum Surelite Laser 102 996 0207 This page is intentionally blank...