Section 5: Maintenance
996-0207
67
11. With a cotton swab wet each end of the exposed flashlamp electrode with water, and then using a
gentle twisting motion, remove the white plastic lamp retainers.
12. Pull firmly on either flashlamp electrode and push on the other to remove the lamp.
13.
Do not
touch the glass barrel of the new lamp. Hold the lamp by the metal electrodes. Inspect
glass barrel and clean with lens tissue and methanol if necessary. Note anode/cathode orientation
of new lamp and insert the lamp into the head according to “ +” and “ -” stickers on the side of
the head.
14. Slide the new lamp into the laser head and install the Teflon backing ring and then the black “ O”
ring over the ends of the lamp electrodes.
Note
:
Wet all parts with distilled water before reinsertion.
15. Press black lamp “ O” rings flush to laser head end plates before installing nylon lamp retainers.
16. Wet the exposed electrode and then replace the white plastic lamp retainers. Use a gentle pushing,
twisting motion.
17. Once the lamp retainers have been slipped over the lamp electrodes, check to make sure the lamp
is centered in the head and then bolt the lamp retainers in place. Tighten screws for snug fit. Do
not over tighten.
18. Place the laser head in base cradle while checking orientation once again and begin to screw the
thumbwheel counter clockwise to reattach the laser head to its cradle. Gently rock the laser head
as you tighten the thumbwheel. The head should mate firmly against the cooling water “ O”
rings.
19. Reattach the lamp connectors at each end. Make sure lamp electrodes are dry before sliding lamp
connectors over the lamp ends.
20. Reattach the two leads to the thermal switch.
21. Reconnect head plug cable at rear of the laser bench (large green/black military style connector).
22. Flow test the lamps. To do this plug in the ac power cord and then perform steps 1 and 2 in the
Daily Startup Procedure.
23. If the lamp leaks, repeat this procedure from step 2. If the laser head leaks from around the
cradle tighten the thumbwheel screw a bit more. If it continues to leak remove the head and
inspect cradle “ O” rings for damage.
24. Log the date and action in “ Operator Notes” at the end of manual and the procedure is complete.
Replacing DI filter
N
ECESSARY
E
QUIPMENT
•
New DI filter (PN#313-0099)
•
1 ½ gallons distilled water (6 liters)
•
3/32" Allen wrench
Summary of Contents for Surelite EX 10
Page 8: ...Continuum Surelite Laser 6 996 0207 This page is intentionally blank...
Page 12: ...Continuum Surelite Laser 2 996 0207 This page is intentionally blank...
Page 26: ...Continuum Surelite Laser 16 996 0207...
Page 36: ...Continuum Surelite Laser 26 996 0207 This page is intentionally blank...
Page 57: ...Section 3 System Operation 996 0207 47 Figure 27 Surelite chronograph...
Page 80: ...Continuum Surelite Laser 70 996 0207 This page is intentionally blank...
Page 92: ...Continuum Surelite Laser 82 996 0207...
Page 94: ...Continuum Surelite Laser 84 996 0207 This page is intentionally blank...
Page 106: ...Continuum Surelite Laser 96 996 0207 This page is intentionally blank...
Page 107: ...Appendix B Operator Notes...
Page 108: ...Continuum Surelite Laser 98 996 0207...
Page 109: ...Appendix B Operator Notes 996 0207 99...
Page 110: ...Continuum Surelite Laser 100 996 0207...
Page 111: ...Appendix B Operator Notes 996 0207 101...
Page 112: ...Continuum Surelite Laser 102 996 0207 This page is intentionally blank...