Continuum Surelite Laser
34
996-0207
12. Reset voltage using
ARROW UP
button on front of PCU to original setting (noted in
step 7) and return to continuous operation.
13. Locate the SLT/SLF adjust knob on the
side
of laser housing assembly and slowly rotate
knob for maximum energy. The knob should not have to be rotated more than one full
turn either clockwise or counter clockwise. Unless it is suspected to be grossly
misadjusted, do not rotate more than one full turn either clockwise or counter clockwise.
14. Locate SLD adjust knob on
top
of laser housing assembly and slowly rotate knob for
maximum energy. Unless gross misadjustment is suspected,
do not
rotate more than one
full turn in either direction.
15. Keep repeating steps 12 and 13 until the 355 nm energy has been maximized.
Fourth Harmonic Generation (SLF)
Table 4 Surelite fourth harmonic specification
Quadrupling
266 nm
Surelite I
10
Surelite I
20
SLI
30
Surelite II
10
SLI
20
Surelite lll
10
Surelite EX
10
Energy(mJ)
60
45
30
80
60
100
90
Stability (%)
± 7
Pulsewidth
~1 ns less than input of 1,064 nm
Polarization
>95% horiz.
Linewidth(cm-1)
1.0
Performance Parameters at 266 nm:
1. Shut down laser system.
2. Remove housing cover and its cover plate shown on page
32
to allow adjustment knob
of SLT or SLF to seat properly.
3. Locate the second harmonic crystal model SLD-2 and install. See procedure on page
36
.
Note:
For maximum 266 nm energy the second harmonic crystal model SLD-2
must be used with the fourth harmonic crystal model SLF. Using the second
harmonic crystal model SLD-1 will lower 266 nm energy by >25%.
4. Locate the fourth harmonic crystal model SLF and install. See procedure on page
37
.
5. Install 266 nm dichroics in Continuum’s optional Surelite Separation Package (SSP).
6. Position the SSP as indicated in
Figure 13
on page
24
.
7. Perform
Daily Startup
procedure with the single shot mode operational. See page
37
.
8. Note voltage reading on front of PCU and then lower lamp voltage (use
ARROW
DOWN
button) ~200 V to just above lasing threshold.
9. With >OD~4 goggles on, hold a piece of white paper (an index or business card) in
front of the exit port and press the single shot button. A small blue spot should be seen
on the card. The IR will be fluorescent white.
Summary of Contents for Surelite EX 10
Page 8: ...Continuum Surelite Laser 6 996 0207 This page is intentionally blank...
Page 12: ...Continuum Surelite Laser 2 996 0207 This page is intentionally blank...
Page 26: ...Continuum Surelite Laser 16 996 0207...
Page 36: ...Continuum Surelite Laser 26 996 0207 This page is intentionally blank...
Page 57: ...Section 3 System Operation 996 0207 47 Figure 27 Surelite chronograph...
Page 80: ...Continuum Surelite Laser 70 996 0207 This page is intentionally blank...
Page 92: ...Continuum Surelite Laser 82 996 0207...
Page 94: ...Continuum Surelite Laser 84 996 0207 This page is intentionally blank...
Page 106: ...Continuum Surelite Laser 96 996 0207 This page is intentionally blank...
Page 107: ...Appendix B Operator Notes...
Page 108: ...Continuum Surelite Laser 98 996 0207...
Page 109: ...Appendix B Operator Notes 996 0207 99...
Page 110: ...Continuum Surelite Laser 100 996 0207...
Page 111: ...Appendix B Operator Notes 996 0207 101...
Page 112: ...Continuum Surelite Laser 102 996 0207 This page is intentionally blank...