89
-
Message
Cause
Alarm Level
IS1 TOO HIGH
SYS measuring value of channel 1 is above upper alarm limit.
User-selectable
IS1 TOO LOW
SYS measuring value of channel 1 is below lower alarm limit.
User-selectable
ID1 TOO HIGH
DIA measuring value of channel 1 is above upper alarm limit.
User-selectable
ID1 TOO LOW
DIA measuring value of channel 1 is below lower alarm limit.
User-selectable
IM1 TOO HIGH
MAP measuring value of channel 1 is above upper alarm limit.
User-selectable
IM1 TOO LOW
MAP measuring value of channel 1 is below lower alarm limit.
User-selectable
IS2 TOO HIGH
SYS measuring value of channel 2 is above upper alarm limit.
User-selectable
IS2 TOO LOW
SYS measuring value of channel 2 is below lower alarm limit.
User-selectable
ID2 TOO HIGH
DIA measuring value of channel 2 is above upper alarm limit.
User-selectable
ID2 TOO LOW
DIA measuring value of channel 2 is below lower alarm limit.
User-selectable
IM2 TOO HIGH
MAP measuring value of channel 2 is above upper alarm limit.
User-selectable
IM2 TOO LOW
MAP measuring value of channel 2 is below lower alarm limit.
User-selectable
Technical alarms:
Message
Cause
Alarm
Level
Remedy
IBP1 SENSOR OFF
IBP cable of channel 1
falis off from monitor.
LOW
Make sure that cable is properly
connected.
IBP2 SENSOR OFF
IBP cable of channel 2
fails off from monitor.
LOW
Make sure that cable is properly
connected.
IBP(1,2) INIT ERR
IBP module failure
HIGH
Stop using measuring function of IBP
module, notify biomedical engineer or
Our service staff.
IBP(1,2) INIT ERR1
IBP(1,2) INIT ERR2
IBP(1,2) INIT ERR3
IBP(1,2) INIT ERR4
IBP(1,2) INIT ERR5
IBP(1,2) INIT ERR6
IBP(1,2) INIT ERR7
IBP(1,2) INIT ERR8
IBP(1,2) COMM STOP
IBP(1,2) module failure
or communication failure
HIGH
Stop using measuring function of IBP
module, notify biomedical engineer or
Our service staff.
IBP(1,2) COMM ERR
IBP(1,2) communication
error
HIGH
Stop using measuring function of IBP
module, notify biomedical engineer or
Our service staff.
IBP1 ALM LMT ERR
Functional safety failure
HIGH
Stop using measuring function of IBP
module, notify biomedical engineer or
Our service staff.
IBP2 ALM LMT ERR
Functional safety failure
HIGH
Stop using measuring function of IBP
module, notify biomedical engineer or
Our service staff.
Prompt message (general alerts):
Message
Cause
Alarm Level