Available options are 12.5mm/s, 25.0 mm/s.
■
AVG TIME
4s, 8s, 16s represent times that SpO
2
average value is counted.
■
DEFAULT:
Pick this item to access the SpO
2
DEFAULT CONFIG dialog box, in which you can select FACTORY DEFAULT
CONFIG or USER DEFAULT CONFIG. After selecting one item and exiting the dialog box, the system will pop up the
dialog box asking for the your confirmation.
12.6 Alarm Description and Prompt
SpO
2
Alarm Message
When the alarm switches are set to ON in relevant menus, the physiological alarms caused by the parameter exceeding
the alarm limit may possibly trigger the recorder to automatically output alarming parameter value and corresponding
waveforms.
Tables below describe the possible physiological alarms, technical alarms and prompt messages occurring during SpO
2
measurement.
Physiological alarm:
Message
Cause
Alarm Level
SpO
2
TOO HIGH
SpO
2
measuring value is above upper alarm limit.
User-selectable
SpO
2
TOO LOW
SpO
2
measuring value is below lower alarm limit.
User-selectable
PR TOO HIGH
PR measuring value is above upper alarm limit.
User-selectable
PR TOO LOW
PR measuring value is below lower alarm limit.
User-selectable
Technical alarms:
Message
Cause
Alarm Level
Remedy
SpO
2
SENSOR OFF
SpO
2
sensor may be
disconnected from the
patient or the monitor.
LOW
Make sure that the monitor and the
patient are in correct connection with
the cables.
SpO
2
INIT ERR
SpO
2
module failure
HIGH
Stop using the measuring function of
SpO
2
module, notify biomedical
engineer or our service staff.
SpO
2
INIT ERR 1
SpO
2
INIT ERR 2
SpO
2
INIT ERR 3
SpO
2
INIT ERR 4
SpO
2
INIT ERR 5
SpO
2
INIT ERR 6
SpO
2
INIT ERR 7
SpO
2
INIT ERR 8
SpO
2
COMM STOP
SpO
2
module failure or
communication error
HIGH
Stop using the measuring function of
SpO
2
module, notify biomedical
engineer or our service staff.
SpO
2
COMM ERR
SpO
2
module failure or
communication error
HIGH
Stop using the measuring function of
SpO
2
module, notify biomedical
engineer or our service staff.
SpO
2
ALM LMT ERR
Functional safety failure
HIGH
Stop using the measuring function of
SpO
2
module, notify biomedical
engineer or our service staff.
PR ALM LMT ERR
Functional safety failure
HIGH
Stop using the measuring function of
SpO
2
module, notify biomedical
engineer or our service staff.
Prompt message (include general alerts):
Message
Cause
Alarm Level
SpO
2
EXCEED
SpO
2
measuring value exceeds the range.
HIGH
PR EXCEED
PR measuring value exceeds the range.
HIGH
SEARCH PULSE
SpO
2
module is searching for pulse.
No alarm