87
-
Figure
Errore. L'origine riferimento non è stata trovata.
-9 IBP Calibration Menu
Calibrate the transducer:
Turn the knob to select the item CH1 CAL VALUE, press and turn the knob to select the pressure value to be calibrated
for channel 1. Then turn the knob to select the item CALIBRATE to start calibrating channel 1.
Turn the knob to select the item CH2 CAL VALUE, press and turn the knob to select the pressure value to be calibrated
for channel 2. Then turn the knob to select the item CALIBRATE to start calibrating channel 2.
The pressure calibration of the monitor
Figure
Errore. L'origine riferimento non è stata trovata.
-10 IBP Calibration
Caution
Mercury calibration should be performed by the biomedical engineering department either whenever a new
transducer is used, or as frequently as dictated by your Hospital Procedures Policy.
The purpose of the calibration is to ensure that the system gives you accurate measurements.
Before starting a mercury calibration, a zero procedure must be performed.
If you need to perform this procedure yourself you will need the following pieces of equipment:
1
.
Standard sphygmomanometer
2
.
3-way stopcock
3
.
Tubing approximately 25 cm long
The Calibration Procedure: (SEE Figure
Errore. L'origine riferimento non è stata trovata.
-10)
Warning
You must never perform this procedure while patient is being monitored.