![Contec Medical Systems Co. CMS8000 User Manual Download Page 124](http://html1.mh-extra.com/html/contec-medical-systems-co/cms8000/cms8000_user-manual_2656215124.webp)
-
117
-
before measuring in IBP1
"IBP2 NEED ZERO-CAL"
Zero calibrating must be done
before measuring in IBP2
Do zero calibrating for IBP2
"TB SENSOR OFF"
TB sensor is not connected
correctly.
Check the connection of TB sensor.
"ECG NOISE"
Rather large interference signals
appear in the ECG signals.
Check the connection of ECG lead wire.
Check the current situation of the patient.
Check if the patient moves a lot.
"XX INIT ERR X"
XX has error X during
initialization.
Re-start up the monitor or re-plug in/out the
module. If the error still exists, contact the
manufacturer.
"XX COMM STOP"
XX cannot communicate with the
host.
"XX COMM ERR"
XX cannot communicate normally
with the host.
XX represents all the parameter modules in the system such as ECG, NIBP, SpO
2
, IBP module, etc.
"XX ALM LMT ERR"
The alarm limit of XX parameter
is modified by chance.
Contact the manufacturer for repair.
"XX RANGE EXCEEDED"
The measured value of XX
parameter has exceeded the
measuring range of the system.
Contact the manufacturer for repair.
XX represents the parameter name in the system such as HR, ST1, ST2, RR, SpO
2
, IBP, NIBP, etc.
"CO
2
Sensor Faulty"
The Sensor Source Current Failure
Check that the sensor is properly
plugged in. Reinsert or reseat the
sensor if necessary. If error
persists,return sensor to factory for
servicing.
"CO
2
Sensor Over temp"
The sensor temperature is greater than
40
℃
Make sure sensor is not exposed to
extreme heat. If error persists, return
sensor to factory for servicing.
"CO
2
Check Sampling Line"
This error occurs whenever the
pneumatic pressure is outside the
expected rang
Check that the sampling line is not
occluded or kinked.
"CO
2
Zero Error"
An error was found during Zero
To clear, check airway adapter and
clean if necessary. If this does not
correct the error, perform an adapter
zero.
"CO
2
Out of Range"
The value being calculated is greater
than the upper CO
2
limit.
If error persists, perform a zero.
"CO
2
Check Airway Adapter"
Usually caused when the airway adapter
is removed from the sensor or when
there is an optical blockage on the
windows of the airway adapter. May
also be caused by failure to perform
Zero to When adapter type is changed.
To clear, clean airway adapter if
mucus or moisture is seen. If the
adapter is clean, perform a Zero.
"CO
2
not initialized"
Barometric Pressure or gas
compensatins have not been set since
power on.
Set the Barometric Pressure and gas
compensations to clear this error.
"REAL CLOCK NEEDSET"
When the system displays 2000-1-1, the
system gives this prompt reminding the
user that the current system time is not
right.
Re-set up the system time. It is better
to set up the time just after the start-up
and prior to monitoring the patient.
After modifying the time, the user had
better re-start up the monitor to avoid
storing error time.
"REAL CLOCK NOT EXIST"
The system has no cell battery or the
battery has run out of the capacity.
Install or replace the rechargeable
battery.