57
-
displayed in the alarm area of the monitor while technical alarms and prompt messages unable to trigger alarms are
then displayed in the information area of the monitor. This section does not describe the content about Arr. and ST
analysis.
Among physiological alarms, those belonging to the type that the parameter has exceeded the limits may activate the
recorder to automatically output the parameters and related measured waveforms when the alarms occur on the
condition that the alarm record switch in the related menu is On.
Tables below describe respectively the possible various alarms those may occur during the measurement.
Physiological alarms:
Message
Cause
Alarm level
ECG LOST
No ECG signal of the patient is detected.
HIGH
HR TOO HIGH
HR measuring value is above the upper alarm limit
User-selectable
HR TOO LOW
HR measuring value is below the lower alarm limit
User-selectable
Technical alarms:
Message
Cause
Alarm
level
Remedy
ECG LEAD OFF
ECG electrodes fall off the
skin or ECG cables fall off
the monitor.
LOW
Make sure that all electrodes, leads
and patient cables are properly
connected.
ECG V LEAD OFF or
ECG C LEAD OFF
ECG LL LEAD OFF or
ECG F LEAD OFF
ECG LA LEAD OFF or
ECG L LEAD OFF
ECG RA LEAD OFF or
ECG R LEAD OFF
ECG INIT ERR
ECG module failure
HIGH
Stop using measuring function
provided by ECG module, notifies
biomedical engineer or Our service
staff.
ECG INIT ERR1
ECG INIT ERR2
ECG INIT ERR3
ECG INIT ERR4
ECG INIT ERR5
ECG INIT ERR6
ECG INIT ERR7
ECG INIT ERR8
ECG COMM STOP
Occasional communication
failure
HIGH
If failure persists, notify biomedical
engineer or Our service staff.
ECG COMM ERR
Occasional communication
failure
HIGH
If failure persists, notify biomedical
engineer or Our service staff.
HR ALM LMT ERR
Functional safety failure
HIGH
Stop using HR alarm function,
notify biomedical engineer or Our
service staff.
ECG NOISE
ECG measuring signal is
greatly interfered.
LOW
Make sure the patient is quiet, the
electrodes are properly connected
and AC power system is well
grounded.
Prompt messages (include general alerts):
Message
Cause
Alarm Level
HR EXCEED
HR measuring value exceeds the measurement range.
HIGH
11.7 ST Segment Monitoring
11.7.1 ST Segment Monitoring
■
The ST algorithm has been tested for accuracy of the ST segment data. The significance of the ST segment