TEMP alarm limits:
Max. TEMP HI
Min. TEMP LO
Step
T1, T2
50
0
0.1
TD
50
0
0.1
UNIT
To set temperature unit (ºC or ºF).
DEFAULT
Pick this item to access the TEMP DEFAULT CONFIG dialog box, in which the user may
select whether the FACTORY DEFAULT CONFIG or the USER DEFAULT CONFIG is to
be used. After selecting any of the items and exiting the dialog box, the system will pop up
the dialog box asking for the user’s confirmation.
14.3 TEMP Alarm message
Among physiological alarms, those belonging to the type that the parameter has exceeded the limits may activate the
recorder to automatically output the parameters and related measured waveforms when the alarms occur on the
condition that the alarm record switch in the related menu is On.
Tables below describe the possible physiological alarms, technical alarms and prompt messages occurring during
TEMP measurement.
Physiological alarms:
Message
Cause
Alarm Level
T1 TOO HIGH
Measuring value of channel 1 is above upper alarm limit.
User-selectable
TI TOO LOW
Measuring value of channel 1 is below lower alarm limit.
User-selectable
T2 TOO HIGH
Measuring value of channel 2 is above upper alarm limit.
User-selectable
T2 TOO LOW
Measuring value of channel 2 is below lower alarm limit.
User-selectable
TD TOO HIGH
Difference between two channels is larger than upper limit.
User-selectable
Technical alarms:
Alarm Message
Cause
Alarm
Level
Remedy
T1 SENSOR OFF
Temperature cable of channel 1
may be disconnected from the
monitor.
LOW
Make sure that the cable is properly
connected.
T2 SENSOR OFF
Temperature cable of channel 2
may be disconnected from the
monitor.
LOW
Make sure that the cable is properly
connected.
T1 ALM LMT ERR
Functional safety failure
HIGH
Stop using alarming function of
TEMP module, notify biomedical
engineer or Our service staff.
T2 ALM LMT ERR
Functional safety failure
HIGH
Stop using alarming function of
TEMP module, notify biomedical
engineer or Our service staff.
TD ALM LMT ERR
Functional safety failure
HIGH
Stop using alarming function of
TEMP module, notify biomedical
engineer or Our service staff.
Prompt message:
Message
Cause
Alarm Level
T1 EXCEED
Measuring value of channel 1 is beyond measuring range.
HIGH
T2 EXCEED
Measuring value of channel 2 is beyond measuring range.
HIGH
14.4 Care and Cleaning
Warning
Before cleaning the monitor or the probe, make sure that the equipment is switched off and disconnected
from the power line.