26
-
Medium
Alarm indicator flashes in yellow with low frequency.
Low
Alarm indicator lights on in yellow.
Alarm Sound
The high/medium/low-level alarms are indicated by the system in following different audio ways:
Alarm level
Audio prompt
High
Mode is “DO-DO-DO------DO-DO, DO-DO-DO------DO-DO”, which is triggered
once every 8 seconds.
Medium
Mode is “DO-DO-DO”, which is triggered once every 8 seconds.
Low
Mode is “DO-”, which is triggered once every 8 seconds.
Note
When alarms of different levels occur at the same time, the monitor prompts the one of the highest level.
Alarm Setup
The setup of the alarms can be realized in the alarm menu.
Press the “ALARM SETUP” button on the SYSTEM SETUP menu to call up “ALARM SETUP” menu (default menu)
as shown below.
Figure 4-1 Alarm Setup
ALARM VOL: which has 7 selections: 1~7.
ALM REC TIME: which has three selections: 8S, 16S, 32S.
ALM PAUSE TIME: refers to the alarm suspension time span, which has three selections: 1MIN, 2MIN, 3MIN.
ALM TYPE:UNLATCH.UNLATCH refers to the situation that once the alarm condition is discharged, the alarm
will disappear automatically.
KEY VOL:there are 8 options:off,1~7.
4.2 Alarm Verification during Power On
During the monitor power on, audible and visual alarm capability will be tested by the system.
Every time when the monitor powers on, alarm beeps “DO-”, and the LED indicator on the display device flashes
orange once. If no beeps heard or no alarm indicator flashing viewed, do not use this device to monitor any patient, and
notify Customer Service Center.
4.3 Alarm Cause
Alarm occurs when:
1.Physiological alarm is evoked;
2. Alarm for error of the system (technical alarm) is evoked;
3. General alert occurs.
A. Conditions that activate the parameter alarms:
When the measurement value exceeds the alarm limit and the alarm is set to“ON”. Alarm will not activate if the alarm
is set to “OFF”.
B. Conditions that activate the system alarms (technical alarm):
Upon the system error, the monitor prompts alarm immediately and proceeds corresponding remedy, stops all
monitoring and eliminates the final results in order to avoid faulted treatment. If more than one error occur, they will be
displayed by turns.
C. General alert