85
-
ALM REC: Select “ON” to enable recording during the IBP alarm or to OFF to disable the alarm recording
function.
SWEEP: used to select the scanning speed of the IBP wave. Two selections are available: 12.5 mm/s or 25 mm/s.
IBP1 UNIT/IBP2 UNIT: used to select the pressure unit (mmHg / kPa / cmH2O).
FILTER:non filter,smooth,normal.
ALM LIMIT SET: used to access the sub-menu of IBP ALM LIMIT SET, in which the user may set up the upper
and lower alarm limit of systolic pressure, diastolic pressure and mean pressure respectively for channel 1 and
channel 2.
SCALE ADJUST: used to access the sub-menu of IBP PRESS RULER ADJUST, in which the user may adjust the
position of the high, reference and low scales for the two waveforms displayed on the screen.
EXPAND PRESSURE: used to access the sub-menu of IBP EXPAND PRESS SET, in which the user may select
the pressure name to be represented by P1, P2.
DEFAULT: pick this item to access the IBP DEFAULT CONFIG dialog box, in which the user may select whether
the FACTORY DEFAULT CONFIG or the USER DEFAULT CONFIG is to be used. After selecting any of the
items and exiting the dialog box, the system will pop up the dialog box asking for the user’s confirmation.
EXIT: used to exit the menu and return to the main screen.
Warning
Before set the alarm limits, confirm to choose the correct label.
Figure
Errore. L'origine riferimento non è stata trovata.
-7 IBP ALM LIMIT SETUP
The alarm occurs when the value exceeds the set limits.
IBP alarm limits:
Pressure Label
Max. Alarm High
(mmHg)
Min. Alarm Low
(mmHg)
Step
(mmHg)
ART
300
-10
1
PA
300
-10
1
CVP
300
-10
1
RAP
300
-10
1
LAP
300
-10
1
ICP
300
-10
1
P1
300
-10
1
P2
300
-10
1