![Contec Medical Systems Co. CMS8000 User Manual Download Page 44](http://html1.mh-extra.com/html/contec-medical-systems-co/cms8000/cms8000_user-manual_2656215044.webp)
-
37
-
The user may select the start time of review in the item of START.
Then the user may select the end time of review. Two selections are available: current time and the user-defined time.
For user-defined end time, the user can use the knob to select.
2.
ALARM RECALL EVENT
In the pull-down list of ALARM RECALL EVENT, the user can select the parameter whose alarm events he wants to
review. The selections include ALL(alarm events of all parameters), ECG, RESP, SpO
2
, NIBP, IBP,TEMP,CO
2
.
After setting up all the review conditions, press the “ALARM RECALL” button to access “ALARM RECALL”
window.
ALARM RECALL
The ALARM RECALL window is as shown below, in which following data are displayed:
①
Time span (Format: month-day-year hour: minute- month-day-year hour: minute).
②
Event type.
③
Serial number (Format: NO. xx of XX ).
④
The value at the time of alarm. NIBP result is with time.
⑤
Two 8/16/32-second waveforms.
Figure 7-5 Alarm Recall Menu
To view all waveforms during the alarming process
Pick L-RIGHT and turn the knob to view all 8/16/32-second waveforms stored.
To view other alarm events
Events of up to 71 are listed chronologically from the latest to the earliest. Pick UP-DOWN button and turn the knob to
view later or earlier events.
Recording
Pick REC to print our all data and waveform of this event.
7.5 SD Operate
The user can review patient data on the monitor or on PC by SD card. Here only introduce reviewing method on the
monitor.
Prepare an empty SD card which capacity is at least 2G. The SD card mounted on the monitor can savetrend data
respectively for HR,PVCs,ST1,SpO
2
,PR,RR,T1,T2 ,TD and 72-hour waveform of ECG. The trend data's resolution is 1
minute.
Note
Please first set the patient's information correctly before inserting SD CARD.
Need to save different patient's data in one SD CARD, you should unmount SD CARD successfully, and
then modify patient's information. Make sure that Patient No. is different.
①
②
④
③
⑤