![Contec Medical Systems Co. CMS8000 User Manual Download Page 39](http://html1.mh-extra.com/html/contec-medical-systems-co/cms8000/cms8000_user-manual_2656215039.webp)
-
32
-
menu and Pick ”WAVE” button to access “ARR WAVE RECALL”
menu. Then press “REC” button to output the Arr. Waveform and related
information currently displayed on the screen.
Alarm review recording
Access the “ALM RECALL” window from “ALARM RECALL TIME”
menu from “SYSTEM MENU” and pick “REC” button to print out the
alarm review waveform and related information currently displayed in
the “ALARM RECALL” window.
NIBP review recording
Access the “NIBP RECALL” window from “SYSTEM MENU” and
pick “REC” button to print out the NIBP information currently displayed
in the window.
Titration table recording
Access the “DRUG CALC” menu from the “SYSTEM MENU” menu.
Pick the “TITRATION” button in the menu to access the “TITRATION”
window. Pick the “REC” button to print out the titration currently
displayed in the window.
Note
You can press REC/STOP button on the recorder to stop the current recording process.
Access the “RECORD” menu from the “SYSTEM SETUP” menu. Then pick the “CLEAR REC TASK”
button to stop all recording tasks, and clear all store of alarm.
6.4 Recorder Operations and Status Messages
Record Paper Requirement
Only standard 50 (+0/-1) mm thermosensitive record paper can be used, otherwise the recorder may not function, the
recording quality may be poor, and the thermosensitive printhead may be damaged.
Function Properly
■
When the recorder is working, the record paper goes out steadily. Do not pull the paper, or the recorder will be
damaged.
■
Do not operate the recorder without record paper.
Paper Out
When "RECORDER OUT OF PAPER" alarm is displayed, the recorder cannot start. Please insert record paper
properly.
Inserting Paper
■
Open the recorder catch.
■
Insert a new roll of paper into the paper cassette, printing side facing the thermosensitive printhead.
■
Give out the paper from the recorder outlet.
■
Close the recorder catch.
Note
Be careful when inserting paper. Avoid damaging the thermosensitive printhead. Unless when inserting
paper or shooting troubles, do not leave the recorder catch open.
Removing Paper Jam.
When the recorder functions or sounds improperly, open the recorder catch to check for a paper jam. If the
paper jam only re-insert the paper.