![Contec Medical Systems Co. CMS8000 User Manual Download Page 85](http://html1.mh-extra.com/html/contec-medical-systems-co/cms8000/cms8000_user-manual_2656215085.webp)
-
78
-
RANGE
EXCEEDED
Measuring range exceeds the
specified upper limit.
HIGH
Reset NIBP module, if failure persists,
stop using measuring function of NIBP
module, notify biomedical engineer or
Our service staff.
EXCESSIVE
MOTION
Affected by arm motion, signal
noise is too large or pulse rate is
not regular.
LOW
Make sure that the patient under
monitoring is motionless.
OVER
PRESSURE
Pressure has exceeded the
specified upper safety limit.
HIGH
Measure again, if failure persists, stop
using measuring function of NIBP module
and notify biomedical engineer or Our
service staff.
SIGNAL
STURATED
Excessive motion
LOW
Stop the patient from moving.
PNEUMATIC
LEAK
During pneumatic test, leak is
detected.
LOW
Check and replace the leaking parts, if
required, notify biomedical engineer or
Our service staff.
NIBP SYSTEM
FAILURE
Operation of blood pressure
pump system is failed.
HIGH
Stop using measuring function of NIBP
module, notify biomedical engineer or
Our service staff.
CUFF TYPE ERR
Cuff type does not comply with
the patient type.
LOW
Select appropriate cuff type
NIBP TIME OUT
Measuring time has exceeded
120 seconds (adult) or 90
seconds (neonatal).
HIGH
Measure again or use other measuring
method.
NIBP
ILLEGALLY
RESET
Abnormal module reset
HIGH
Reset again
MEASURE FAIL
Problem happens when
measuring the curve. The system
cannot perform measurement,
analysis or calculation.
HIGH
Check the cuff. Make sure that the patient
under monitoring is motionless. Measure
again.
Prompt message: (display in the prompt area below NIBP value)
Message
Cause
Alarm Level
Manual measure…
During manual measuring mode.
No alarm
Cont measuring…
During continuous measuring mode.
Auto measuring…
During automatic measuring mode.
Please start
After selecting interval between measurements in MENU
Measurement over
Press START/STOP key during measuring to stop measurement
Calibrating…
During calibrating
Calibration over
Calibration over
Pneum testing…
During pneumatic test
Pneum test over
pneumatic test over
Resetting...
NIBP module in resetting
Reset failed
NIBP module reset failed
13.5 Maintenance and Cleaning
Warning
Do not squeeze the rubber tube on the cuff.
Do not allow liquid to enter the connector socket at the front of the monitor.
Do not wipe the inner part of the connector socket when cleaning the monitor.