118
-
"SYSTEM WD FAILURE"
The system has serious error.
Re-start up the system. If the failure
still exists, contact the manufacturer.
"SYSTEM SOFTWARE ERR"
"SYSTEM CMOS FULL"
"SYSTEM CMOS ERR"
"SYSTEM EPGA FAILURE"
"SYSTEM FAILURE2"
"SYSTEM FAILURE3"
"SYSTEM FAILURE4"
"SYSTEM FAILURE5"
"SYSTEM FAILURE6"
"SYSTEM FAILURE7"
"SYSTEM FAILURE8"
"SYSTEM FAILURE9"
"SYSTEM FAILURE10"
"SYSTEM FAILURE11"
"SYSTEM FAILURE12"
"KEYBOARD NOT
AVAILABLE";
The keys on the keyboard cannot be
used.
Check the keys to see whether it is
pressed manually or by other object. If
the key is not pressed abnormally,
contact the manufacturer for repair.
"KEYBOARD COMM ERR";
The keyboard has failure, which cannot
be used.
Contact the manufacturer for repair.
"KEBOARD ERROR";
"KEYBOARD ERR1";
"KEYBOARD ERR2";
"NET INIT ERR(G.)"
The network part in the system has
failure. The system cannot be linked to
the net.
Contact the manufacturer for repair.
"NET INIT ERR(Ram)"
"NET INIT ERR(Reg)"
"NET INIT ERR(Mii)"
"NET INIT ERR(Loop)"
"NET ERR(Run1)"
"NET ERR(Run2)"
"NET ERR(Run3)"
"5V TOO HIGH"
The power part of the system has
failure.
If the prompt appears repeatedly,
contact the manufacturer for repair.
"5V TOO LOW"
"POWER ERR3"
"POWER ERR4"
"12V TOO HIGH"
"12V TOO LOW"
"POWER ERR7"
"POWER ERR8"
"3.3V TOO HIGH"
"3.3V TOO LOW"
"CELL BAT TOO HIGH"
Cell battery has problem.
Replace the battery. If the failure still
exists, contact the manufacturer.
"CELL BAT TOO LOW"
The cell battery has low capacity or the
cell battery is not installed or the
connection is loose.
"RECORDER SELFTEST ERR"
During the selftest, the system fails
connecting with the recorder module.
Execute ‘Clear Record Task’ function
in the recorder setup menu to
re-connect the host and the recorder. If
the failure still exists, contact the
manufacturer for repair.
"RECORDER VLT HIGH"
The recorder module has voltage failure.
Contact the manufacturer for repair.
"RECORDER VLT LOW"