84
-
to be infused.
4.
Position the transducer so that it is at the same level with the patient’s heart, approximately mid-axillary line.
5.
Check if you have selected the correct label name. See the next section for details.
6.
Zero the transducer. See the next section for details.
Figure
Errore. L'origine riferimento non è stata trovata.
-4 IBP Monitoring
15.4 IBP Menu
Pick the IBP hot key on the screen to access the IBP SELECT menu shown as following:
Figure
Errore. L'origine riferimento non è stata trovata.
-5 IBP SELECT Menu
Pick the IBP SETUP item to call up the IBP SETUP menu as following:
Figure
Errore. L'origine riferimento non è stata trovata.
-6 IBP SETUP Menu
The items to be set up in the menu include:
ALM: Select “ON” to enable alarm prompt and data storage during IBP alarm. Select “OFF” to disable audio
alarm and prompt the
symbol beside “IBP” numeric.
ALM LEV: used to set up the alarm level. Three levels are available: HI, MED, LO.