113
-
Initialization Time
Mainstream
Capnogram displayed in less than 15 seconds at an ambient
temperature of 25
℃
, full specifications within 2 minutes.
Sidestream
Capnogram displayed in less than 20 seconds at an ambient
temperature of 25
℃
, full specifications within 2 minutes.
Mainstream Rise Time
Less than 60ms-Adult Reusable or Single-Patient-Use Airway Adapter
Less than 60ms-Infant Reusable or Single-Patient-Use Airway Adapter
Actualization interval
about 1 Sec
Sidestream Delay Time:
2~3Sec
Alarm range
CO
2
0~150 mmHg
InsCO
2
0~150 mmHg
AwRR
2 ~ 150 BPM
Suffocation Alarm Delay
AwRR
10~60 Sec.