Long
480 hrs, 1 Min. Resolution
Alarm Event Recall
71 alarm events of all parameters and 8/16/32seconds of corresponding waveform.
NIBP Measurement Recall
At least 4800 NIBP measurement data.
SD card
72 hrs ECG waveform
TREND review
2.7 ECG
Lead Mode
5 Leads ( R, L, F, N, C or RA, LA, LL, RL, V)
Lead selection
I, II, III, aVR, aVL, aVF, V,
Waveform
2 ch
Lead mode
3 Leads ( R, L, F or RA, LA, LL)
Lead selection
I, II, III,
Waveform
1 ch
Gain
2.5mm/mV, 5.0mm/mV, 10mm/mV, 20mm/mV, 40mm/mV
Scan speed
12.5mm/s, 25mm/s, 50mm/s
HR
Measure and Alarm Range
Adult
15 ~ 300 bpm
Neonate/Pediatric
15 ~ 350 bpm
Accuracy
±1% or ±1bpm, which great
Resolution
1 bpm
Sensitivity
> 200 uV P-P
Differential Input Impedance
> 5 MΩ
CMRR
Monitor
≥ 100dB
Operation
≥ 100 dB
Diagnosis
≥60dB
Electrode offset potential
±300mV
Leakage Current
< 10 μA
Baseline Recovery
≤ 5 s After Defi.
Input bias current (lead off detection)
≤ 0.1 μA
(
driver lead ≤ 1 μA
)
ECG Signal Range
±8 mV (Vp-p)
Notch
50Hz/60Hz(Notch filter can be turned on or off manually)
Bandwidth
Surgery
1 ~ 20 Hz(+0.4dB,-3dB)
Monitor
0.5Hz
~
40Hz(+0.4dB,-3dB)
Diagnostic
0.05Hz~75Hz(+0.4dB,-3dB);76Hz~150Hz(+0.4dB,-4.5dB)
Calibration Signal
1 mV (Vp-p), ±5% Accuracy
ST Segment Monitoring Range
Measure and Alarm
-2.0 ~ +2.0 mV
Accuracy
-0.8 ~ +0.8mV
0.04mV or
10% which is greater
Other unspecified
ARR Detecting
Type
ASYSTOLE, VFIB/VTAC, COUPLET, BIGEMINY, TRIGEMINY,
RONT, VT>2, PVC, TACHY, BRADY, MISSED BEATS, PNP, PNC
Alarm
Available
Review
Available
Tall T-wave rejection capability
1.2mV
Heart rate averaging
the average value of the latest 6 R-R intervals which
have ignored the maximum and minimum
Updating rate of the display
1s
Heart rate meter accuracy and response to irregular rhythm:
Bigeminy ventricular
80±1bpm
Bigeminy ventricular alternative lente
60±1bpm
Bigeminy ventricular alternative rapid
120±1bpm
Systoles bidirectional
90±1bpm
Response time to heart rate meter to change in heart rate